top of page

CAMAG - World Leader in Planar Chromatography

 CAMAG > HPTLC Pro > Module Development

CAMAG HPTLC Pro Module Development

The HPTLC PRO Module DEVELOPMENT is a part of the CAMAG® HPTLC PRO SYSTEM, which is the first fully automated HPTLC system worldwide.

​

The HPTLC PRO Module DEVELOPMENT revolutionizes the development step. With the new chamber geometry, this step is performed much faster. The gas phase can be actively circulated during development (conditioning) using a loop system through the conditioning tube, which contains a solvent system of choice. This procedure allows to further optimize the separation of the target analytes. By using the same solvent system as the developing solvent, the gas phase is regenerated.

​

The CAMAG® HPTLC Software visionCATS allows you to prepare several analysis files, which can be used sequentially. In this mode, the autonomous development of up to five different HPTLC glass plates (20 x 10 cm) with up to three different mobile phases is supported.

​

As a part of the CAMAG® HPTLC PRO SYSTEM, the Module DEVELOPMENT shares several components with the other modules and has a similar design. Each module can be operated either as stand-alone (Module Mode) or within the HPTLC PRO SYSTEM (System Mode). For the System Mode the modules are connected and a conveyor transports the HPTLC plate from one module to the other.

Key features

​

  • Part of the fully automated HPTLC PRO SYSTEM

  • Development of up to 5 HPTLC plates (20 x 10 cm)

  • Supports up to 3 different developing solvents

  • Chamber geometry streamlined to speed up the development

  • Gas phase can be actively circulated during development

  • HPTLC glass plates (20 x 10 cm), only

  • Optimal cleaning between different developing solvents

  • Software-controlled by visionCATS

Technical Specifications​

​

  • Nitrogen or clean air pressure: 5 – 8 bar (70 – 110 psi)

  • Operating temperature: 15 – 30 °C
  • Recommended working temperature: 20 – 25 °C

  • Humidity: Maximum relative humidity 80 % for temperatures up to 30 °C 

  • Plate types: HPTLC glass plates 20 x 10 cm, layer thickness 200 µm

  • Dimensions (W x D x H): 384 x 550 x 510 mm

  • Weight ~ 41 kg

Ordering Information

​

ORL-060-3000

​

CAMAG® HPTLC PRO Module DEVELOPMENT

can be operated either as stand-alone or as part of the fully automated CAMAG® HPTLC PRO SYSTEM. Allows to autonomously develop up to five different HPTLC glass plates (20 x 10 cm) with up to three different mobile phases. The gas phase can be actively circulated.

​

Includes:

  • 2 Carrier for HPTLC glass plates (20 x 10 cm)

  • Set of Bottles for start-up.

bottom of page